Instructions for Use
9 Analyzing Exam Data and Creating Reports
2660021169042 Rev. A 2018-03
9.5 Using Presets
2660021169042 Rev. A 2018-03
149 / 246
9.5.5 Creating Custom Presets
There are four presets reserved for you to define your own custom
presets.
You can create one or two of each type of custom perset:
•
Create Custom(Local)
: for an individual patient.
•
Create Custom(Global)
: for all patients.
The algorithm you select determines how to calculate the bound-
aries:
•
Sum
computes the
summation
of
all flow pixel values
for each
A-line (axial direction) between the layers.
•
Max
computes the
average
of the
highest 5 flow pixel values
for each A-line (axial direction) between layers.
To Create a Custom Preset:
Prerequisite
þ
Logged in to review station (or instrument): Logging In as
Operator or Data Analyst [
}
73]
þ
The scan report or analysis is open: Opening a Report or
Analysis [
}
120]
Action
1. To create a preset you can use for all patients, click on an
unused preset labeled
Create Custom (Global)
.
2. To create a preset for this patient, click on an unused preset
labeled
Create Custom (Local)
.
ð
Custom Enface Settings
opens.
3. For
Name
, type a name for the preset.
Summary of Contents for PLEX Elite 9000
Page 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Page 54: ...Empty page for your notes...
Page 106: ...Empty page for your notes...
Page 174: ...Empty page for your notes...
Page 182: ...Empty page for your notes...
Page 210: ...Empty page for your notes...
Page 218: ...Empty page for your notes...
Page 224: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 228: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 245: ...Empty page for your notes...