Instructions for Use
1 Safety and Certifications
2660021169042 Rev. A 2018-03
1.3 Safety
2660021169042 Rev. A 2018-03
19 / 246
1.3.4 Laser Safety and Compliance
Figure 1: Laser Classification Label
Figure 2: Laser Compliance Label
Complies with 21 CFR Subchapter J and with FDA performance
standards for laser products except for deviations pursuant to Laser
Notice No. 50, dated June 24, 2007.
1.3.5 Electrical Safety
Class I Equipment
- Protection against electrical shock.
1.3.6 Printer and Peripherals Safety
WARNING!
Non-compliance with system requirements laid out in
standard IEC 60601–1 for instruments externally connected
to non-medical peripheral devices (i.e. printers, non-system
storage devices, etc.)
compromises patient safety.
u
If a non-medical peripheral device is located within 1.5 m from
the patient, the usage of an isolation transformer is required.
u
If the peripheral device is located outside the patient
environment (beyond 1.5 m) and is connected to the
instrument, a separation device must be used or there shall be
no electrical connection between the non-medical peripheral
device and the instrument.
WARNING!
Powering peripherals directly through a wall socket
could result in electrical shock to the patient and/or examiner.
u
When using a printer in the USB configuration, always power
the printer through an isolation transformer. Some ZEISS
equipment comes with an isolation transformer that may be
Summary of Contents for PLEX Elite 9000
Page 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Page 54: ...Empty page for your notes...
Page 106: ...Empty page for your notes...
Page 174: ...Empty page for your notes...
Page 182: ...Empty page for your notes...
Page 210: ...Empty page for your notes...
Page 218: ...Empty page for your notes...
Page 224: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 228: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 245: ...Empty page for your notes...