Instructions for Use
15 Specifications
2660021169042 Rev. A 2018-03
15.1 Imaging Parameters
2660021169042 Rev. A 2018-03
211 / 246
15 Specifications
15.1 Imaging Parameters
15.1.1 OCT and OCT Angiography Imaging
Methodology
Swept Source OCT
Optical Source
Swept Source Tunable Laser
Center Wavelength
Centered between 1040 nm-1060 nm;
Sweep range: 980 nm – 1120 nm
Scan Angle
10° minimum
Scan Speed
100,000 A-Scans/sec
Operating Beam Power
(nominal)
4.5 mW at cornea
A-Scan Depth
3.0 mm (in tissue), 1536 points
Axial Resolution (optical)
6.3 µm (in tissue)
Axial Resolution (digital)
1.95 µm (in tissue)
Transverse Resolution*
20 µm
Field of View
56°
Minimum Pupil Diameter
2.5 mm
*Transverse (Lateral) resolution is calculated from beam size at the pupil.