![background image](http://html.mh-extra.com/html/zeiss/mmz-m/mmz-m_operating-instructions-manual_927572085.webp)
Mounting / changing the probe
61211-1270302
MMZ M
6-11
3
Cover the adapter plate receptacle
[2]
.
4
Loosen the screw; width across flats: 2.5 mm
[3]
.
The screw is located on the right-hand side of the ram.
5
Remove the VAST XT or VAST probe from the dovetail guide of the
ram
[4]
.
6
Slide the RDS articulating system onto the dovetail guide (
without
RDS adapter plate) –
[5]
.
7
Tighten the screw to 2.2 Nm
[6]
.
8
Reactivate the drives.
The new probing system is automatically recognized. However, it must
be initialized in the measuring software. Then the styli of the stylus sys-
tem must be qualified and the workpiece must be aligned again during
the computer-controlled manual run. Please refer to the operating in-
structions for the measuring software.
Summary of Contents for MMZ M
Page 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Page 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Page 74: ...Installation 5 6 61211 1270302 MMZ M ...
Page 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Page 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Page 146: ...2Glossary 2 61211 1270302 MMZ M ...
Page 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Page 153: ......
Page 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...