![background image](http://html.mh-extra.com/html/zeiss/mmz-m/mmz-m_operating-instructions-manual_927572037.webp)
Safety on the CMM
61211-1270302
MMZ M
2-19
> 900
> 900
> 400
> 800
7
8
10
4
5
6
1
2
3
> 800
(1)
> 1000
> 1000
> 1000
(14)
(13)
11
13
12
14
15
9
16
Examples of blocking off a danger zone
1
Distance between CMM and wall, minimum 800 mm
2
Wall
3
Light barrier receiver
4
Laser beam
5
CMM, top view
6
Light barrier transmitter
7
Distance between controller cabinet and CMM, minimum 1000 mm
8
Controller cabinet
9
Distance between compressed air connection and laser beam, minimum
1000 mm
10
Control console
11
Signal tower
12
Distance between signal tower and laser beam, minimum 1000 mm
13
Deflecting mirror
14
Distance between measuring range and laser beam, minimum 900 mm
15
Distance between CMM and laser beam, minimum 400 mm
16
Measuring range