Evaluation of the measuring results
7-24
61211-1270302
MMZ M
Source of error
Cause
Workpiece condition
–
The workpiece shows signs of deformation and roughness.
–
The workpiece is soiled.
–
The workpiece is covered with oil, for example drops of oil in the
small bores.
–
The workpiece material is too soft.
Magnetic field
–
A malfunction was caused by a magnetic field.
Possible causes of a magnetic field: for example workpiece, clamp-
ing tool, stylus system extensions.
Qualification
–
Different axis clamping set than for qualification.
–
Different measuring force set than for qualification (depending on
the measuring method, with measuring probes)
Probing
–
Probing has not been performed perpendicular to the probing sur-
face.