![background image](http://html.mh-extra.com/html/zeiss/mmz-m/mmz-m_operating-instructions-manual_927572138.webp)
Care
9-12
61211-1270302
MMZ M
2 Adhesive label indicating replacement or cleaning of the filter mats
3 Air filters of the fans
1
Remove the air filter cover.
2
Clean or replace the filter mat.
The air filter lamella must lie deeper on the outside than on the in-
side. See illustration.
3
Place the cover on the air filter.
1
2
Alignment of the cover
1 Cover
2 Lamella of the cover
Checking the compressed air filters
The service unit is equipped with at least two filters. The CMM is usually
equipped with a fine filter and an extra fine filter. The filters have to be
checked regularly. The transparent filter caps make it possible to check
the degree of soiling.
1
2
(2)
Service unit for compressed air supply
1 Fine filter with filter element
2 Extra fine filter with filter element
Summary of Contents for MMZ M