Preparation for measuring operation
61211-1270302
MMZ M
7-15
NOTE
For measuring probing systems, a
Tensor qualification
should be per-
formed. This is important because the measuring force causes a certain
amount of stylus bending. Tensor qualification makes it possible to cor-
rect the bending for all subsequent probing operations.
Tensor qualification must be carried out
by means of vectors
. All axes
can move.
NOTE
Different qualification methods are available depending on the applica-
tion. Please refer to the operating instructions for the measuring soft-
ware.
1
Install the stylus system.
2
The menu for the qualification appears in the measuring software.
Please refer to the operating instructions for the measuring software.
3
On the control console, select the stylus you want to qualify.
4
Probe the reference sphere in the shaft direction of the stylus and
measure its diameter.
The shaft of the reference sphere may not be probed during the
qualification.
Probing directions during qualification
A Correct
B Wrong
1 Stylus shaft
2 Probing direction
5
Check the result.
Summary of Contents for MMZ M
Page 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Page 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Page 74: ...Installation 5 6 61211 1270302 MMZ M ...
Page 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Page 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Page 146: ...2Glossary 2 61211 1270302 MMZ M ...
Page 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Page 153: ......
Page 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...