2.2 Power
Requirements
☞
The LSM 510 comes with a mains power supply cord and plug, either CEE red (230 V, 16 A,
3 phases), or CEE yellow (115 V, 32 A, 3 phases), and with the matching mains socket outlet.
Line voltage
230 V AC: 220...240 V AC (±10 %)
115 V AC: 100...125 V AC (±10 %)
Line frequency
50...60 Hz
50...60 Hz
LSM incl. VIS laser
−
Max. current
2 phases at 16 A
Phase 1 = 1.8 kVA max.
Phase 2 = 2 kVA max.
2 phases at 25 A
Phase 1 = 1.8 kVA max.
Phase 2 = 2 kVA max.
−
Power consumption
2000 VA per phase
2000 VA per phase
−
Power plug
CEE red (230 V, 16 A):
3 N+PE, phases 1 and 2
connected
CEE yellow (115 V, 32 A):
3 N+PE, phases 1 and 2
connected
Argon UV laser
- Line Voltage
208...240 V AC
(±10 %) 50 / 60 Hz
208...240 VAC
(±10 %) 50 / 60 Hz
−
Max. current
1 phase at 63 A
Note: For Line Voltage 220 V the
connector and power plug are rated
for 63 Amps, However wiring and
fuse should be rated for 32 Amps.
1 phase at:
208 V: 34 Amps
230 V: 31 Amps
240 V: 29 Amps
−
Power consumption
7000 VA
7000 VA
Class of protection
I
I
Type of protection
IP 20
IP 20
Overvoltage category
II
II
Pollution degree
2
2
☞
If the line voltage in your country is 115 V AC, you need to order an additional 2.5 kW step-
up-transformer, part no. 234.366, to be able to run the ArKr laser. Reason: The ArKr laser
requires a 220 V input.
Summary of Contents for LSM 510 META
Page 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Operating Manual Release 3 2...
Page 8: ...INTRODUCTION LSM 510 Carl Zeiss LSM 510 META VIII B 45 0008 e 10 02...
Page 10: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss Contents LSM 510 META 1 2 B 45 0008 e 10 02...
Page 20: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss LSM 510 META 1 12 B 45 0008 e 10 02...
Page 22: ...LSM 510 SETUP REQUIREMENTS LSM 510 Carl Zeiss Contents LSM 510 META 2 2 B 45 0008 e 10 02...
Page 52: ...QUICKSTART LSM 510 Carl Zeiss Contents LSM 510 META 4 2 B 45 0008 e 10 02...