LSM 510
QUICKSTART
LSM 510 META
Quick Start in the Expert Mode
Carl Zeiss
B 45-0008 e 10/02
4-15
4.3.2.2
Axiovert 200 M
•
Click on the
Acquire
button in the toolbar of the
Main
menu.
•
Click on the
VIS
button in the
Acquire
subordinate toolbar.
•
Click on the
Micro
button.
−
The
Axiovert Control
window appears on
the screen.
•
Put the specimen on the stage - make sure the
specimen is mounted securely and flat. Use the
supplied Convallaria specimen at first.
You can view the specimen in either fluorescence
(reflected light) or transmitted light.
•
To view the specimen in transmitted light, set
the
Reflector Turret
position to
None
and
activate the
Transmitted Light
panel by
clicking on the
Transmitted Light
button.
Select the
On
button and control the intensity
by the slider.
•
Select an objective with low magnification by
clicking in the
Objective
panel of the
Axiovert
Control
window.
•
Set the microscope to KÖHLER illumination
manually
(see
Axiovert 200 M
operating
manual).
•
Select the specimen area to be examined by
moving the XY-stage and focus exactly on the
selected area.
•
Close the
Axiovert Control
window. Click on
the
LSM
button.
4.3.2.3
Axioskop 2 FS
MOT
For setting the Axioskop 2 FS
MOT
, proceed in the same way as described for Axioplan 2 imaging
MOT
.
Since the Axioskop 2 FS
MOT
is not motorized (except for the z motor drive), all microscope settings have
to be made manually.
The
Micro
button in the
Acquire
subordinate toolbar of the
Main
menu is therefore not available.
Fig. 4-13
Axiovert Control window
Summary of Contents for LSM 510 META
Page 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Operating Manual Release 3 2...
Page 8: ...INTRODUCTION LSM 510 Carl Zeiss LSM 510 META VIII B 45 0008 e 10 02...
Page 10: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss Contents LSM 510 META 1 2 B 45 0008 e 10 02...
Page 20: ...NOTES ON DEVICE SAFETY LSM 510 Carl Zeiss LSM 510 META 1 12 B 45 0008 e 10 02...
Page 22: ...LSM 510 SETUP REQUIREMENTS LSM 510 Carl Zeiss Contents LSM 510 META 2 2 B 45 0008 e 10 02...
Page 52: ...QUICKSTART LSM 510 Carl Zeiss Contents LSM 510 META 4 2 B 45 0008 e 10 02...