Select the appropriate filters and activate the 
channels. 

• 

Click the 

Excitation

 button to select the laser 

lines and set the attenuation values 
(transmission in %) in the displayed window.  

For the configuration of the beam path, please 
refer to the application-specific configurations 
depending on the used dyes and markers and the 
existing instrument configuration. 
 

 

• 

Clicking on the 

Spectr 

button 

opens the 

Detection Spectra & 

Laser Lines …

 window (Fig. 13) to 

display the activated laser lines for 
excitation (colored vertical lines) and 
channels (colored horizontal bars). 

 

 

• 

Clicking on the

 Config 

button 

opens the 

Track Configurations

 

window (Fig. 14) to load, store or 
delete track configurations. 

 

• 

For storing a new track configuration enter a 
desired name in the first line of the 

Configurations

 list box and click an 

Store

• 

For loading an existing configuration select it in 
the list box and click on 

Apply

• 

For deleting an existing configuration select it in 
the list box and click on 

Delete

 
 

Setting for multi track configuration in 
Channel Mode 

The 

Multi Track

 function permit several tracks to 

be defined as one configuration (

Channel Mode 

Configuration

) for the scan procedure, to be 

stored under any name, reloaded or deleted.  

The maximum of four tracks with up to 8 channels 
can be defined simultaneously and then scanned 
one after the other. Each track is a separate unit 
and can be configured independently of the other 
tracks with regard to channels, Acousto-Optical 
Tunable Filters (AOTF), emission filters and dichroic 
beam splitters.  

• 

Select 

Channel Mode

 if necessary (Fig. 15). 

 

Fig. 13 

Detection Spectra & Laser Lines … 
window  

 

Fig. 14 

Track Configurations window  

 

Fig. 15 

Configuration Control window for 
Multi Track 

Summary of Contents for LSM 510 Inverted

Page 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 3 5 July 2005...

Page 2: ...e LSM 510 and LSM 510 META laser scanning microscope including its original accessories and compatible accessories from other manufacturers may only be used for the purposes and microscopy techniques...

Page 3: ...nents switch also to ON Now the complete system is ready to be initialized with the LSM Software Switching on the HBO 100 mercury lamp Switch on the HBO 100 mercury lamp via the switch of the power su...

Page 4: ...10 Switchboard window Clicking on this button activates the complete LSM hardware on line mode Click on the Start Expert Mode button in the LSM 510 Switchboard window The LSM 510 Expert Mode Main menu...

Page 5: ...ate toolbar Click on the New button in the File subordinate toolbar The Create New Database window appears Select drive C or D from pull down menu Create a new directory if needed Fig 6 Create New Dat...

Page 6: ...cular tube of the microscope Click on the VIS button to set the microscope for direct observation via the eyepieces of the binocular tube lasers are off Click on the TV button to set the microscope ca...

Page 7: ...ting the microscope for reflected light Click on the Reflected Light button to open the shutter of the HBO 100 mercury lamp Click on the Reflector button and select the desired filter set by clicking...

Page 8: ...ick on the Single Track button in the Configuration Control window Fig 12 Click on the Descanned button if necessary The Beam Path and Channel Assignment panel of the Configuration Control window disp...

Page 9: ...he first line of the Configurations list box and click an Store For loading an existing configuration select it in the list box and click on Apply For deleting an existing configuration select it in t...

Page 10: ...position and diameter When Line button is selected the same rules apply as for Frame Fast Settings Add Track button An additional track is added to the configuration list The maximum of four tracks c...

Page 11: ...lts in the best signal to noise ratio Scan speed 8 usually produces good results Us speed 6 or 7 for superior images Choosing the dynamic range Select the dynamic range 8 or 12 Bit per pixel in the Pi...

Page 12: ...same Optical Size This is important for colocalization studies Image acquisition Once you have set up your parameter as defined in the above section you can acquire a frame image of your specimen Use...

Page 13: ...gh it appears blue on the screen Blue zero minimum Adjusting the laser intensity Set the Pinhole to 1 Airy Unit Fig 21 Set the Detector Gain high When the image is saturated reduce AOTF transmission i...

Page 14: ...upper position of the specimen area where the Z Stack is to start Click on the Mark First button to set the upper position of the Z Stack Then focus on the lower specimen area where the recording of t...

Page 15: ...ing off the system Click on the File button in the Main menu and then click on the Exit button to leave LSM 5 software program Fig 5 If any lasers are still running you should shut them off now in the...

Page 16: ......

Reviews: