Axiovert 40
B 46-0031 e 11/03
2Ć23
2.8
Replacing the specimen stage
•
Remove the screw cover (2Ć28
ā
/
ā
1
) from the stage surface.
•
Using the SW 3 ball-headed screwdriver, remove three fastening screws (2Ć28
ā
/
ā
2
) and the stage (2Ć28
ā
/
ā
3
).
•
Fit and fasten the selected specimen stage (specimen stage glass (Section 2.8.1) heatable stage or temperĆ
able microscope stage (Section 2.8.2)).
Fig. 2Ć28
Replacing the specimen stage
1
2
2
3