Safety notes for the use of the HBO 50
CAUTION
While in operation, the HBO 50 is under high pressure. It may therefore only be operated in
a closed microscope illuminator.
•
The cooling process of the lamp housing must not be obstructed by covers.
•
Before you can change the lamp, the HBO 50 must cool down for approx. 15 minutes.
•
While in operation, the lamp emits UV radiation. Avoid direct exposure of eyes and skin. It is advisable to
wear protective eyewear when handling the microscope illuminator.
CAUTION
The HBO 50 must be replaced after expiry of its average life of 100 h. The risk of explosion
increases when the average life is exceeded.
The operating time of the HBO 50 can be read from the hour meter.
This warning label on the rear of the illuminator means:
Caution: Hot surface!
Allow the lamp to cool down before touching it.
Replacing the HBO 50 lamp
CAUTION
Switch off power supply and disconĆ
nect the power plug of the lamp from
the socket of the power supply. AlĆ
low lamp and lamp housing to cool
down (for at least 15 min.).
•
Loosen the clamping screw on the microscope
stand to remove the HBO 50 illuminator from the
stand.
•
Place the illuminator on a flat work surface.
•
Using the provided SW 3 ball-headed screwĆ
driver, loosen screw (3Ć4ā/ā
1
) and remove the lamp
housing (3Ć4ā/ā
2
).
•
Depress spring lever (3Ć4ā/ā
5
) and, holding lamp
(3Ć4ā/ā
4
) on heat sink (3Ć4ā/
ā3
), withdraw the lamp
from its holder.
•
Place the heat sink with lamp on the work surface
so as to make the clamping screw on the heat sink
accessible.
Fig. 3Ć4
Replacing the lamp of the HBO 50
1
2
3
4
6
5
Summary of Contents for Axiovert 40
Page 1: ...Operating Manual Axiovert 40 Inverted Microscope...
Page 2: ......
Page 3: ......
Page 11: ...Axiovert 40 B 46 0031 e 11 03 VII GENERAL VIEW Axiovert 40 CFL...
Page 12: ...Axiovert 40 11 03 B 46 0031 e VIII...
Page 16: ...Axiovert 40 11 03 B 46 0031 e XII...
Page 24: ...Axiovert 40 11 03 B 46 0031 e 1 8 1 4 System overview Fig 1 3 1 Axiovert 40 system overview...
Page 25: ...Axiovert 40 B 46 0031 e 11 03 1 9 Fig 1 3 2 Axiovert 40 system overview...
Page 26: ...Axiovert 40 11 03 B 46 0031 e 1 10 Fig 1 3 3 Axiovert 40 system overview...
Page 65: ...Axiovert 40 B 46 0031 e 11 03 A 1 ANNEX Index List of abbreviations...