Changing the 6 V, 30 W Transmitted-
Light Halogen Lamp
The following procedure is required to change the
lamp of the integrated 6 V, 30 W illuminator:
•
Switch off the extern lamp supply; if required,
allow the illuminator to cool down for approx.
15 minutes.
•
Disconnect the line cable from the line.
•
Hold the ventilation grille (4-3/
1
) on its grip and
pull it off in upward direction.
•
Remove the lamp insert by pressing its handle
(4-3/
2
) together.
•
Replace the halogen lamp 6 V, 30 W (4-3/
3
).
☞
Do not touch the lamp bulb with your bare hands; if required, clean the bulb using pure
alcohol
before
switching it on for the first time to prevent dirt from burning in.
•
Make sure that the lamp carrier plate exactly fits the positioning pins.
•
Attach lamp insert again by slightly pressing the handle together
•
Attach ventilation grille again until stop.
4.3
Requesting Service
All repairs of mechanical, optical or electronic components inside the instrument and of the electrical
components of the Axiolab A may only be performed by Carl Zeiss service staff or specially
authorized
personnel.
To ensure the optimum setting and trouble-free function of your microscope even for a longer period of
time, we would recommend you to conclude a service/maintenance contract with Carl Zeiss.
In the case of subsequent orders or when service is required, please get in touch with your local Carl
Zeiss agency.
Fig. 4-3
Changing the 6 V, 30 W
transmitted-light halogen lamp
Summary of Contents for Axiolab A
Page 1: ...Axiolab A Reflected Light Microscope Operating Manual...
Page 16: ...Carl Zeiss Axiolab A 1 6 06 99 B 40 015 e Fig 1 3 Axiolab A microscope configurations...
Page 18: ...Carl Zeiss Axiolab A 1 8 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 1 of 2...
Page 20: ...Carl Zeiss Axiolab A 1 10 06 99 B 40 015 e Fig 1 4 Axiolab A accessories 2 of 2...
Page 26: ...Carl Zeiss Axiolab A 1 16 06 99 B 40 015 e Fig 1 5 Axiolab A function elements 1 of 2...
Page 27: ...Axiolab A Carl Zeiss B 40 015 e 06 99 1 17 Fig 1 5 Axiolab A function elements 2 of 2 1 1 2...
Page 34: ...Carl Zeiss Axiolab A 1 24 06 99 B 40 015 e...
Page 36: ...Carl Zeiss Axiolab A 2 2 B 40 015 e 06 99...
Page 44: ...Carl Zeiss Axiolab A 3 2 B 40 015 e 06 99...
Page 68: ...Carl Zeiss Axiolab A 4 2 B 40 015 e 06 99...