Image Acquisition
Axiocam 105 color Instal Reference Guide
31
Note
:
Please note that a gamma setting of 0.45 is required for correct color
reproduction. This can be achieved in the Live image by activating the
button. Only the display is influenced by this function. The data
remain unchanged.
Image Orientation
An Axiocam 105 color image is adjusted in such a way that the image
orientation corresponds to the image in the eyepiece when mounted on an
upright microscope with the front of the camera facing you.
Setting options:
Original
Vertically reflected
Horizontally reflected
Rotated 180 degrees
This starting position can be adapted via this menu to suit your
requirements. If, for example, the camera’s cable routing needs to be
changed on an inverse microscope, the camera can be mounted at a
different angle, without the correspondence between the image orientation
and the eyepiece being lost.
Note:
Activating an image rotation can reduce the maximum live-image
speed.
Shading Correction
Activate this function if you notice, e.g. with brightfield images, a reduction
in image brightness at the image edge or uneven color representation. For
this calibration process make sure that the camera is viewing a completely
empty image. To ensure that this is the case, move your sample on the
microscope completely out of the field of view.