Testing Alarm Out Status
Alarm outputs that have been disabled will deactivate
the alarm output contact located on
TB1
, terminals
#17
and
#18,
refer to the Wiring Control Document on page
117 in Appendix D
.
You may test the alarm output in
general by choosing the
Alarm SwitchTest
option
below,
figure 168-169
, or you may choose to
Simulate
specific alarms as illustrated in this section as well.
Choose
*Set
in the main menu,
figure 163
.
Choose
*Par
– parameters in the Set Selection menu,
figure 164
.
Choose
*Flow
or
*Time
in the Set Parameters menu,
figure 165
.
Choose
*Alarm
from the Proportional-to-Time or Flow
menus,
figure 166
.
Choose
*Test
from the Alarm Out Status menu,
figure
167.
Choose
*SW
in the Switch / Simulation menu,
figures
168.
To activate the Alarm Test Switch,
figures 169
,
press
and release the
Select
key. The entered value will flash
when it is chosen. Use the
Down Arrow
or
Up Arrow
keys to change the setting.
ON / Relay Open
indicates
the alarm relay is in the test mode.
OFF / Relay Closed
indicates the relay is in the normal mode. Press the
Enter
key to load the entry into memory. When the unit
is in the
Alarm Test Mode
, a bright red light will flash in
the alarm light indicator located just below the
SELECT /
ENTER
switch. The display will stop flashing when the
entered value is loaded into the memory. Press the
Up
Arrow
key to advance to return to the Switch / Simula-
tion
menu.
Choose
*Sim
in the Switch / Simulation menu. Any time
an alarm is simulated or real, a bright red light will flash
in the alarm light indicator located just below the
SE-
LECT / ENTER
switch. Any time a Non-Alarm is simu-
lated or real, a bright green light will flash in the indicator
light indicator located just below the
Down Arrow
key.
Or
pmp bat vmtr sig tnk
Proportional to Flow
*Set *Alarm * Esc
pmp bat vmtr sig tnk
Proportional to Time
*Set *Alarm * Esc
Figure 163
pmp bat vmtr sig tnk
PropFlow Idle
*Strt *Dsp *Set
pmp bat vmtr sig tnk
Set Selection
*Par *Cal *Esc
pmp bat vmtr sig tnk
Set Parameters
*Flow *Time *Esc
pmp bat vmtr sig tnk
Alarm Output Status
*Set *Test *Esc
Figure 164
Figure 165
Figure 166
Figure 167
pmp bat vmtr sig tnk
Switch / Simulation
*Sw *Sim *Esc
pmp bat vmtr sig tnk
Alarm Switch Test
ON / Relay Open
pmp bat vmtr sig tnk
Alarm Switch Test
Alarm Relay Closed
Figure 168
Figure 169
Figure 169