4.4
Displaying Peak Values
Procedure
Displayed Function
Press the
FUNCTION
key for display C to select Vpk (voltage peak value), Apk (current peak value),
or Wpk (power peak value).
Each time you press FUNCTION, the displayed function changes in the order shown below.
Display
Vpk(+)
Vpk(-)
C
Apk(+)
Apk(-)
Wpk(+)
Wpk(-)
• Vpk, Apk, Wpk, Wh±, and Ah± all illuminate twice. The first time is for the positive measurements and the
second time is for the negative measurements.
• The MATH indicator is on the left side of the 7-segment LED display.
• Press SHIFT before pressing FUNCTION to change the displayed function in reverse order.
Input Element (WT332E/WT333E only)
Press the
ELEMENT
key for display C to select which element to display.
The procedure for selecting an input element is the same as the procedure in section 4.1.
The WT310E and WT310EH only have one input element, so there is no input element selection.
Explanation
Displayed Function
Select which measurement function to show on the display.
• Vpk: The peak voltage is displayed. Select U+pk to display the highest voltage and U-pk to display
the lowest voltage.
• Apk: The peak current is displayed. Select I+pk to display the highest current and I-pk to display the
lowest current.
• Wpk: The peak power is displayed. Select P+pk to display the highest power and P-pk to display
the lowest power.
Maximum Displayed Peak Value, Units, and Unit Prefixes
• Maximum displayed value: 99999 (when the number of displayed digits is set to 5)
• Units: V for the peak voltage, A for the peak current, and W for the peak power
• Unit prefix: m, k, or M
Input Element (WT332E/WT333E only)
The elements that you can select vary depending on the model. Check the model when selecting
elements.
• 1, 2, or 3: Display the measured values of element 1, 2, or 3.
• Σ: Hyphens (------) are displayed instead of measured values.