4.2 Displaying
Apparent Powers, Reactive Powers,
and Power Factors
Procedure
Displayed Function
Press the
FUNCTION
key for display A, B, or D to select VA (apparent power), var (reactive power), or
PF (power factor).
Each time you press FUNCTION, the displayed function changes in the order shown below.
VA
var
Display
A
PF
B
PF
D
• The THD indicator is on the left side of the 7-segment LED display.
• Press SHIFT before pressing FUNCTION to change the displayed function in reverse order.
Input Element (WT332E/WT333E only)
Press the
ELEMENT
key of the display that you selected the display function for (display A, B, or D),
and then select an input element.
The procedure for selecting an input element is the same as the procedure in section 4.1.
The WT310E and WT310EH only have one input element, so there is no input element selection.
Explanation
Displayed Function
Select which measurement function to show on the display.
• VA: Apparent power S is displayed.
• var: Reactive power Q is displayed.
• PF: Power factor λ is displayed.
Maximum Displayed Value, Units, and Unit Prefixes
• Maximum displayed apparent power and reactive power: 99999 (when the number of displayed
digits is set to 5)
• Power factor display range: –1.0000 to 1.0000 (when the number of displayed digits is set to 5)
• Units: VA for apparent power, var for reactive power, and no unit for power factors
• Unit prefix: m, k, or M