2-18
IM WT310E-01EN
2.8 Setting the Measurement Period
Procedure
Follow the procedure indicated by the thick lines in the following menu.
Select the measurement period.
(Display B)
1.
2.
3.
(Display C)
4.
5.
Confirm
the setting.
Close menu.
6.
Explanation
Measurement Period
The measurement period is determined by the input signal that is used as the reference (synchronization
source). The measurement period is set within the data update interval between the first point where the
sync source crosses the level-zero point (center of the amplitude) on a rising slope (or falling slope) and
the last point where the sync source crosses the level-zero point on a rising slope (or falling slope).
If there is not more than one rising or falling slope within the data update interval, the entire data
update interval is set as the measurement period.
Synchronization Source
Because the periods of input signal are detected from both voltage and current signals, you can set
the synchronization source to a voltage signal (VoLt) or a current signal (Curr). The default setting is
VoLt for the WT310E and WT310EH and Curr for the WT332E/WT333E.
• VoLt
Priority is given to the detection of the period of the voltage signal, and the voltage signal is used
as the synchronization source. The voltage signal of each element is used as the synchronization
source for the element. When the period of the voltage signal cannot be detected, the current signal
is used as the synchronization source. When the period of the current signal cannot be detected
either, the measurement period is set to the entire duration of the data update interval.
• Curr
Priority is given to the detection of the period of the current signal, and the current signal is used
as the synchronization source. The current signal of each element is used as the synchronization
source for the element. When the period of the current signal cannot be detected, the voltage signal
is used as the synchronization source. When the period of the voltage signal cannot be detected
either, the measurement period is set to the entire duration of the data update interval.
• oFF
Measurement is not performed in sync with the voltage or current signal. Instead, the measurement
period is set to the entire duration of the data update interval.
Note
• The measurement period for determining the numeric data of the peak voltage or peak current (peak
values) is always the entire span of the data update interval.
• For details about setting the measurement period, see appendix 4.
• When the data update interval (see section 2.10) is set to Auto, the measurement period is determined by
the synchronization source set from the menu described in section 2.10.
• For information about the measurement period for measurement functions used in harmonic measurement
(option), see section 6.3.