• On this instrument, S, Q, λ, and Φ are derived through the computation of the measured values of voltage,
current, and active power. Therefore, for distorted signal input, the value obtained on this instrument may
differ from that obtained on other instruments that use a different method.
• For Q, when the current leads the voltage, the Q value is displayed as a negative value; when the current
lags the voltage, the Q value is displayed as a positive value. The value of QΣ may be negative, because
it is calculated from the Q of each element with the signs included.
Appendix 1 Symbols and Determination of Measurement Functions