![YOKOGAWA WT310E User Manual Download Page 77](http://html.mh-extra.com/html/yokogawa/wt310e/wt310e_user-manual_911416077.webp)
5-7
IM WT310E-01EN
Integration
1
2
3
4
5
6
7
8
App
Index
6.
To reset integration, press
SHIFT+STOP
(
RESET).
The STOP key light will turn off, and displays A and C will show hyphens (”------”).
Explanation
Displayed Function
Select which measurement function to show on the display.
• TIME: The elapsed integration time is displayed.
• Wh: The sum of positive and negative watt hours WP is displayed.
• Wh±: The positive and negative watt hours are displayed. Positive watt hour is WP+, and negative
watt hour is WP-.
• Ah: The total ampere hour q is displayed.
• Ah±: The total ampere hour is displayed or the positive and negative ampere hours are displayed.
Positive ampere hour is q+, and negative ampere hour is q-.
• MATH: When the computation function is set to the average active power during integration, the
average active power during integration is displayed.
Maximum Displayed Value, Units, and Unit Prefixes
• Maximum display
Elapsed integration time: 10000
Integrated values: 99999 (999999 when the unit is MWh or MAh) or –99999 when a negative value
is displayed.
• Units: Wh for the integrated power (watt hour) and Ah for the integrated current (ampere hour)
• Unit prefix: m, k, or M (for integrated values)
Display and Resolution of the Elapsed Integration Time
The elapsed integration time can be displayed using up to 9 total digits for hours, minutes, and
seconds. This instrument shows the elapsed integration time in display A. Because the maximum
number of displayed digits in display A is 6, there are times when all the digits in the elapsed
integration time cannot be displayed.
The number of digits in the displayed time varies as indicated below depending on the elapsed
integration time.
Elapsed Integration Time
Display A Indication
Display Resolution
0 to 99 h, 59 m, 59 s
0.00.00 to 99.59.59
1 s
100 h to 9999 h, 59 m, 59 s
100.00 to 9999.59
1 m
10000 h
10000
1 h
Note
• For details about Wh, Wh±, Ah, and Ah±, see section 5.1.
• For details about the average active power during integration, see section 4.5.
• Regardless of the MAX hold feature, this instrument determines and displays the integrated value by
summing the value that is measured at every data update interval.
5.3 Displaying Integrated Values