8-2
IM WT310E-01EN
8.2 Initializing the Settings
Procedure
Follow the procedure indicated by the thick lines in the following menu.
3.
(Display C)
4.
5.
Confirm the setting.
Execute initialization.
Confirm that you want to
initialize the settings.
2.
Select initialization.
(Display B)
1.
Explanation
Initializing Setup Parameters
You can initialize the setup parameters to their factory defaults. This feature is useful when you want to
cancel all of the settings that you have entered or when you want to redo measurement from scratch.
The default settings are shown below.
Item
Default setting
Display A
Display function: V, element: 1
Display B
Display function: A, element: 1
Display C
Display function: W, element: 1
Display D
Display function: PF, element: 1
Number of displayed digits
Hi (5 digits)
Data update interval
0.25 s
When set to
Auto
Synchronization source
WT310E/WT310EH: U1
WT332E/WT333E: I1
Timeout: 1s
Line filter
OFF
Frequency filter
OFF
Measurement synchronization source
WT310E/WT310EH: VoLt (voltage)
WT332E/WT333E: Curr (current)
Measurement range
Auto range
Measurement range skip
OFF
Measurement mode
RMS
Wiring system (WT332E/WT333E only)
1P3W
Hold
OFF
MAX hold
OFF
Scaling
V:1.000, C: 1.000, F: 1.000
Scaling ON/OFF: OFF
External current sensor scaling constant
10.00
Averaging
Averaging type: Lin (moving average), averaging coefficient: 8
Averaging ON/OFF: OFF
Crest factor
3
MATH equation
WT310E/WT310EH: Voltage crest factor
WT332E/WT333E: Efficiency
Frequency
Voltage frequency and current frequency of element 1
Integration
Reset condition, integration mode: manual
Integration timer value: 0 hr, 00 min, 00 s
Harmonic measurement (option)
PLL source: U1, total harmonic distortion equation: iEC, maximum
measured harmonic order: 50
Harmonic measurement data display ON/OFF: OFF
Storage
Interval: 0 hr, 00 min, 00 s, storage ON/OFF: OFF
D/A output (option)
Output items: default normal measurement items, rated integration
time: 1 hr, 00 min, 00 s