2-5
IM WT310E-01EN
Measurement Conditions
1
2
3
4
5
6
7
8
App
Index
WT310EH
3.
(Display C)
1.
2.
Confirm
the setting.
The above options appear when the crest factor has been
set to 3. The following options appear when the crest factor
has been set to 6 or 6A.
Auto, 20, 10, 5, 2.5, 1, 0.5, SEnSor
WT332E/WT333E
3.
(Display C)
1.
2.
Confirm
the setting.
The above options appear when the crest factor has been
set to 3. The following options appear when the crest factor
has been set to 6 or 6A.
Auto, 10, 5, 2.5, 1, 0.5, 0.25, SEnSor
Note
SEnSor only appears when the external current sensor input option is installed. For details about the
procedure when selecting SEnSor, see section 2.4.
Voltage and Current Range in Quick Configuration Mode
1.
Press the
VOLTAGE
or
CURRENT
key.
In quick configuration mode, the VOLTAGE or CURRENT key and the ▼ and ▲ keys illuminate. No display
appears on the menu. The currently set range appears temporarily in display B, for the voltage range, or
display D, for the current range, and then the measured data reappears.
2.3 Setting the Measurement Range When Using Direct Input