
21-9
IM WT18O1E-02EN
21.8 Performing Zero-Level Compensation
This section explains how to perform zero-level compensation.
►
“Zero-Level Compensation (CAL)” in the features guide
Press
SHIFT+SINGLE
(CAL) to execute zero-level compensation.
Note
• This instrument automatically performs zero-level compensation after you change the measurement
range or input filter.
• To make accurate measurements, we recommend that you execute zero-level compensation after
warming up the instrument for at least 30 minutes.
• If the measurement range and input filter remain the same for a long period of time, the zero level may
change due to the changes in the environment. If this happens, we recommend that you execute zero-
level compensation.
• The integration feature includes an auto calibration feature that automatically performs zero-level
compensation.