![YOKOGAWA WN100 User Manual Download Page 46](http://html.mh-extra.com/html/yokogawa/wn100/wn100_user-manual_911396046.webp)
<9. Parameter Summary>
9-4
IM 01W07D01-01EN
Object
ID
Attribute
ID
Label
Description
Default value Handling
(*
1
)
4: WNO
39
Microphone
Temperature
Coeff B
The coefficient of the microphone temperature
calculation.
Refer to Attribute#36.
This value is determined by hardware design.
Please do not rewrite.
0.0
W
40
Microphone
Initial Calibration
Coeff
The microphone calibration coefficient.
Refer to subsection 8.3 “Procedure to change the
calibration coefficient”.
---
W
41
Microphone Drift
Calibration Coeff
The microphone calibration coefficient.
Refer to subsection 8.3.2 “Sensitivity calibration
procedure”.
0
W
42
Temperature
Characteristic
Coeff
The coefficient of the microphone gain
compensation by the microphone temperature.
This value is microphone specific. Please do not
rewrite.
-0.005
W
43
Temperature
Correction
Switch
The ON/OFF switch of the microphone gain
compensation by the microphone temperature.
0: OFF
1: ON
ON
W
44
Audio Self Test
No use
0
R
45
Octave Band
Switch
The execution switch of the 1/3 octave band
calculation.
0: OFF
1: ON
Refer to “Setting the Octave Band Switch” of
subsection 7.3.5.
OFF
W
46
Sensor
Calibration
Method
This parameter shows the executed calibration type
number.
100: Normal calibration
Normal
calibration
R
47
Sensor
Calibration
Location
A universal parameter to store the comment that
describes where the calibration has been done.
Up to 32 characters can be used.
---
W
48
Sensor
Calibration Date
A universal parameter to store the comment that
describes when the calibration has been done.
Up to 32 characters can be used.
---
W
49
Sensor
Calibration Who
A universal parameter to store the comment that
describes who has done the calibration.
Up to 32 characters can be used.
---
W
50
Octave Band
Type
Indicates the Octave Band Type
3: 1/3 octave band
1/3 octave
band
R
51
Lower Octave
Band Freq.
The lowest center frequency of the 1/3 octave band
In Hz.
25.0
R
52
Octave band
Number
The number of 1/3 octave band.
29
R
53
Octave Band
Calculation
Method Rev
The revision number of octave band calculation.
1
R
54
Sensor
Serial Number
Indicates the sensor number of the transmitter.
---
W
55
FPGA Revision The revision number of the FPGA.
---
R
56
Time
Synch Switch
The ON/OFF switch of the time synchronization
function.
0: OFF
1: ON
Refer to “Setting the Time Synch Switch” on
Subsection 7.3.5.
ON
W
57
Time
Synch Status
The status indication of the time synchronization.
0: Not synchronized
1: Synchronized
Refer to “Setting the Time Synch Switch” on
Subsection 7.3.5.
---
R