IM 11M10A01-03E
A-1
Appendix
r
Appendix 1. Sampling Unit
K9424GA Sampling Unit for OX100/102
This sampling unit is designed for the OX100 Series Oxygen Analyzers when they are
used where a sample gas contains a small amount of organic solvents (isopropyl alco-
hol), such as in reflow furnaces.
The sampling unit comprises two membrane filters to eliminate dust and mist, a suction
pump to draw a sample gas, an activated carbon filter to eliminate organic solvents, and
a flowmeter to monitor sample gas flow rates.
Notes
If repair is required, the is equipment will need to be returned to the factory. For repair
or maintenance, contact your Yokogawa representative.
1. Specifications
Pressure at sampling point
: Atmospheric pressure
Pressure at sample outlet
: Atmospheric pressure
Gas temperature
: 50
8
C max.
Organic solvent content
: ppm to a few percentage level of isopropyl alcohol
(IPA) equivalent
Dust content
: 5 mg/Nm
3
max.
Life expectancy of activated carbon: The expected life of the activated carbon filter is
approx. 50 hours when a sample gas contains 500
ppm of isopropyl alcohol and flows at 500 ml/min.
Power supply
: 100 to 240 V AC, 50/60 Hz
Piping connection
: Rc1/4
Operating environment temperature range: 0 to 40
8
C
Life of disk filter
: When specified flow is no longer possible. Replace
as early as possible.
Pump life
: Approximately one year when used eight hours a
day.
Included Accessories
1. Power cord (1)
A1007WD (AC 100 V standard)
A1009WD (AC 200 V by special order)
2. Fuse (2)
A1109EF (rating 1A)
3. Instruction Manual (1)
K9424JY