IM 11M10A01-03E
5-4
5.4
Calibration Kit (Optional)
The calibration kit for the OX102 oxygen analyzer comprises a standard gas cylinder
and pressure regulating valve to supply standard gas for calibration, when necessary.
F1305.EPS
Standard
Gas
OX102 Sensor
K9424JD
(255)
(310)
(90)
[
75
Pressure
Gauge
Gas Vent Tubing
Needle Valve Assembly
Needle Valve Assembly, with Pressure Reducing Valve
and Fixed Throttle
Rc1/8 (Nominal diameter of piping
connection for K9424JD)
Standard Gas Cylinder
Standard Calibration
Gas (6 bottles/box)
K9424JC
Approx. 340
Approx. 180
Unit : mm
Approx. 300
T1301E.EPS
K9424JA
K9424JB
K9424JR
K9424JS
Part Number
Description
Calibration kit: 1000 ppm, 6 cylinders (1L) + needle valve assembly, with flow adjustment
Calibration kit: 1000 ppm, 6 cylinders (1L) + needle valve assembly, without flow adjustment
Calibration kit: 10000 ppm, 6 cylinders (1L) + needle valve assembly, with flow adjustment
Calibration kit: 10000 ppm, 6 cylinders (1L) + needle valve assembly, without flow adjustment
For a system without flow adjustment, adjust the flow rate to 500 ml/min using the needle valve (K9424JD).
Figure 5.4 Calibration Kit
CAUTION
Shelf life of gas cylinder
The shelf life of a calibration gas shall be one year from the manufacturing date speci-
fied on the label of the cylinder. Yokogawa shall not guarantee accuracy of the equip-
ment, if it is calibrated with an expired calibration gas.