IM 11M10A01-03E
6-20
CAL.4
ppm+CAL
ppm+CAL
ERROR
9310
CAL
ENT
CAL
ENT
F6315.EPS
YES
NO
*1
*2
*3
m
.
%
s
ppm
d
%
s
ppm
d
%
s
ppm
d
Continued from CAL.3
[10000 ppm O
2
Calibration]
Indication
Measurement Mode
m
.
*
( ) CAL4 is for 0-10000 ppm
range calibration. Assuming
the O
2
concentration of
calibration gas is 9310 ppm,
enter this value using
and keys.
Stop supplying calibration gas No.5
Supplying calibration gas No.6 (O
2
concentration
allowance: 8500-9500 ppm)
Calibration value
Entering calibration value
Calculating
Changing
value
Calibration
accepted
Calibration
completed
CAL5 (to the following page)
*1 Entered to skip calibration
*2 Entered to enter calibration value
*3 Entered to return to measurement mode