IM 11M10A01-03E
6-12
(4) Configuring the 1-5 V output signal
This mode is to configure the 1-5 V output signal to correspond to a fixed range.
Configure the signals following the procedure shown in the flow chart below.
This output signal is always set to correspond to a fixed range. Therefore, even if
automatic range switching is selected, this output signal should be configured for each
range.
m
.
ppm+CAL
.
+ 4 - 20mA
Out . 1 - 4
4-20mA
ENT
( * )
F6309E.EPS
( * )
%
d
ppm
s
%
s
ppm
s
( * ) Use [
m
] and [
.
] keys to configure Out.1 to 4.
Out.1 : Full scale 25% O
2
Out 2 : Full scale 10000 ppm O
2
Out 3 : Full scale 1000 ppm O
2
Out 4 : Full scale 100 ppm O
2
Indication
Operational Flow Chart
Measured
Value
Changing
value
Output
accepted
Figure 6.7 Flow Chart of 1-5 V Output Signal Setting