IM 11M10A01-03E
6-11
6. Operation
(3) Configuring the 4-20 mA output signal
This mode is to configure the 4-20 mA output signal for the fixed range. Configure the
signals following the procedure shown in the flow chart below.
This mode is automatically determined if the automatic/manual range switching mode is
set to autorange mode.
m
.
ppm+CAL
4-20mA
Out . 1 - 4
4-20mA
ENT
F6308E.EPS
%
d
ppm
s
%
s
ppm
s
Measured
Value
Changing
value
Output
accepted
(*2)
(*1)
(*2)
Operational Flow Chart
(*1)
When AUTO is selected, the measurement mode remains unchanged.
(*2)
Use [
m
] and [
.
] keys to configure Out.1 to 4.
Out.1 : Full scale 25% O
2
Out 2 : Full scale 10000 ppm O
2
Out 3 : Full scale 1000 ppm O
2
Out 4 : Full scale 100 ppm O
2
Indication
Figure 6.6 Flow Chart of 4-20 mA Output Signal Setting