IM 11M10A01-03E
5-1
5. Component Functions
5.
Component Functions
This chapter describes functions of main equipment of the OX102 oxygen analyzer.
5.1
Sensor
The sensor uses a zirconia current limiting system, enabling direct insertion of the sensor
at elevated temperatures up to 150
8
C. 0 to 250
8
C within 50 mm of sensor tip. The sensor
should be used at a measuring point pressure of 1013
6
40 hPa abs. When replacing the
sensor, enter the data in the Sensor Parameter Table (attached to the sensor control
block) into the converter. This ensures sensor compatibility.
F501.EPS
A
B
B
A
A
B
A
B
1.015
1.043
-65
1.002
-25
0.990
-19
1.429
-17
Sensoe Parameter
T
able
Sensor NO. XXXX
Sensor Parameter Table *
Gain 1
Gain 2
Gain 3
Gain 4
Gain 5
Figure 5.1 Sensor
T5101.EPS
B -17
* Sensor Parameter Table (Example)
Sensor Parameter Table
Sensor No. YA-0007
Gain 1
When replacing the sensor, enter values in the
Sensor Parameter Table into the converter.
See Section 6.3 (1) "Setting Sensor Parameter"
for procedure.
1.015
Gain 2
A 1.043
B -65
Gain 3
A 1.002
B -25
Gain 4
A 0.990
B -19
Gain 5
A 1.429