<9. TROUBLESHOOTING>
9-2
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
Alarm
Number
Name
Explanation
Remedy
NE107
Default
NE107
Changed
X543
Light source failure
Light source intensity
is abnormal.
Check light source
monitoring current and
Lightning IF. If value is
abnormal, open the front
cover and check the cable
connection or assemble
condition. If the cause is
not found, please replace
the Light source part.
S
Enable
X54E
Auto Zero cal result
error
Result of auto zero
calibration is out of
range.
Confirm the solenoid valve
runs correctly by manual
operation. Check if the
zero water flows normally.
If no error is found in the
sampling, check whether
no condensation is found
on Measurement window.
S
Enable
X54F
Auto Cal stability time
limit
Detection value is not
stable during auto
calibration.
S
Enable
Note : X of Alarm Number denotes Channel of the measurement electrode unit.
1: sensor connection number 1-1
5: sensor connection number 2-1
6: sensor connection number 2-2 (not required on the first FD)
Table 9.2
TB820D Analyzer Error Message List (Wash)
Alarm
Number
Name
Explanation
Remedy
Default
(NE107)
Error
config.
(HMI)
009E
Auto wash config. error
Auto wash is not
properly set.
Check auto wash setting.
Converter setting > Auto
wash > Auto calib. setting
C
Enable
009F
Disable auto update of
next start
Auto wash/
calibration is
overdue.
Set the first calibration
start time. Converter
setting > Auto wash> Auto
calib. setting
C
Enable