<9. TROUBLESHOOTING>
9-1
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
9. TROUBLESHOOTING
9.1
When an error occurs
Table 9.1
TB820D Error List
Alarm
Number
Name
Explanation
Remedy
NE107
Default
NE107
Changed
X500
EEPROM error
The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X501
User param. read error
The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X502
Factory param. read
error
The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X503
AD converter failure
AD converter of the
main board failure
The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X504
RAM failure
RAM failure
The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X505
Flash Memory failure
Flash Memory failure The TB820D has become
defective. Contact
Yokogawa service.
F
Disable
X50A
Sensor communication
error
No sensor is
detected.
Check the power supply
connection.
Check the connection
between the detection unit
and the converter.
Check whether the
connected sensor type
matches the one in the
Model and Suffix Code.
F
Enable
X520
Wash/Cal. box error
Relay box does not
work.
Check the connection of
Relay box or configuration
(Enable/Disable.)
S
Disable
X521
Humidity sensor failure Humidity sensor does
not work.
Open the front cover
and check the cable
connection of the front
detector board.
If the cable is correctly
connected, replace the
detector board. When
you replace the detector
board, you need to perform
turbidity calibration.
S
Enable
X540
Turbidity High
Turbidity value is
exceeded high limit
setpoint.
Check measurement
value.
Confirm Turbidity warning
high limit on sensor setting.
S
Enable
X541
Turbidity Low
Turbidity value is less
than low limit setpoint.
Check measurement
value.
Confirm Turbidity warning
low limit on Sensor setting.
S
Enable
X542
Light detection error
Photodetector current
is out of range.
Check the scattered light
current. Identify where
the error occurs : Flow
cell, light source board,
detector board. After then,
bring it back to a normal
state.
S
Enable
Summary of Contents for TB820D