<2. INSTALLATION, PIPING, AND WIRING>
2-1
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
2. INSTALLATION, PIPING, AND WIRING
n
Installation Location
The TB820D turbidimeter should be installed in a location:
• that is protected from direct sunlight and rain, i.e., the instrument should be installed indoors
or in a cubicle;
• that is subject to minimal mechanical vibration;
• where no corrosive gases are present;
• that is not exposed to high temperature and humidity. The temperature should be in the
range of -5 to 55°C, preferably at or around normal temperatures, and its fluctuation small.
The humidity should be kept between 10 to 90%RH. Prevent condensation which may
occur if the sample temperature is lower than the ambient temperature. Also, take protective
measures to prevent a sample and the tap water from freezing, if necessary;
• that provides adequate maintenance space and easy access for maintenance work; and,
• where the drain is provided.
n
Installation Dimensions
Install the TB820D’s detector unit in pipes or walls using their special mounting brackets,
respectively. Note that these mounting brackets are supplied only when specified.
l
Pipe Mounting (Option Code “/U”)
For information on the converter, read FLXA402T IM.
54
141
128
Unit: mm
Mounting 2-inch pipe
(ø60.5)
l
Wall Mounting (Install the detector with four M5 screws)
For information on the converter, read FLXA402T IM.
100
2-R3
2-R3
160
Unit: mm