<2. INSTALLATION, PIPING, AND WIRING>
2-3
IM 12B01B30-02EN 2nd Edition : June 09 , 2021-00
(201)L
(202)N
(002)G
U1
U2
U3
S
L
N
G
(32)NC
(31)C
(33)NO
S1
(42)NC
(41)C
(43)NO
S2
(72)NC
(71)C
(73)NO
S4
(61) +
(62) -
mA1
(65) +
(66) -
mA2
(81) +
(82) -
mA3
(85) +
(86) -
mA4
(21)
(22)
(24)
SHLD(63)
(87) +
(88) -
SHLD(89)
A+(91)
B-(92)
GND(93)
SHLD(94)
Termination(95)
(52)NC
(51)C
(53)NO
S3
Power
2nd input
(-TB, -P1, -C1)
1st input
(82)SHLD
(86)GND
(83)B-
(84)A+
(87)V5
L(101)
N(102)
G(001)
Power
Ultra Sonic
Converter
RELAY
SENCOM
Internal
US terminal
L(301)
N(302)
G(003)
L1
L2
G
(12)
(13)
L(101)
N(102)
G(001)
Power
L(201)
N(202)
G(002)
Detector
L(301)
N(302)
G(003)
Converter
(21)
(22)
Wash
Measure
(31)
(32)
Zero
(41)
(42)
Drain
CN1
SV1 *4
SV2 *4
SV3 *4
SV4 *4
*2
*2
*3
*3
TB820D-
□□
-
□□
-A5
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
Analog output (-N2, -N4)
Analog output (-N2, -N4)
Analog output (-N4)
Analog output (-N4)
Contact input (-N2, -N4)
Analog input (-N4)
RS485 (-R)
Ethernet (-E)
Contact output
(-WR)
TB820D
Right Angle Scattered Light Turbidity Detector
PG400
Pulse Generator for Clean Unit
Power
supply
*1
Power supply cable (C)
Relay box
*1: Power terminal "G" on Relay box must be grounded (ground resistance: 100 ohm or less).
In case of selecting -A5, power supply cable connects with L(101), N(102) and G(001) in Relay box.
*2: Connection cables are 1 m in length normally.
They are available with /L02, /L03, /L05, /L10 or /L20 depending on the length you need.
*3: When suffix code -U1 is specified, PG400 should be purchased separately.
Neither Ultrasonic oscillator cables nor power cable are supplied with the product.
Purchase them separately. See GS 19C01B05-01EN.
*4: In case of selecting -A5, purchase the solenoid valve corresponding input power supply and frequency.
See
●
Solenoid valve (SV1 to SV4).
Cable for SV (F)
Power cable for FLXA402T (B)
SENCOM cable for Turbidity and chlorine (A)
Cable for Relay box control
Power cable for Smart unit (D)
Power cable for PG400
Oscillator cable
Figure 2.2
Wiring TB820D-□□-□□-A5
CAUTION
Turn off power supply to TB820D before wiring. Power rating must comply with TB820D
specifications. Power voltage must match with the one indicated on the name plate.