<3. PREPARATION for OPERATION>
3-2
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
3.4
Filling zero water and Test operation
Supply zero water to the flow cell. When a zero filter is used, first run tap water through the filter
to remove the air inside. After the air is removed enough, keep running the water for more than
one hour.
3.5
Setting Operation Parameters
TB820D is preset with defaults prior to factory shipment.
Before measurement, verify that these factory default settings meet the operating conditions. If
necessary, reconfigure parameters to enable the expected operations. For details of setting data,
When user changes the operation parameter, note down the changed data. Set the appropriate
parameters to suit your operation conditions.
1: Display/Output setting on the converter
See FLXA402T Operation of Converter
2: Setting detector unit parameter. See “4. SETTING PARAMETERS”.
3: Setting Auto wash/Auto calibration
Configure sequence operation by setting up the converter. see
3.6
Calibration
Operate TB820D for about one hour before calibration. Refer to “7. CALIBRATION” for
calibration requirements.