< 7. Calibration >
7-1
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
7.
CALIBRATION
This chapter describes types of calibration enabled with TB820D. Periodical calibration and an
optimization of calibration parameters for Zero or Slope ensure accurate measurement. Zero
shift or the sensitivity correction can also correct the standard and grab sample sensitivity. Clean
the Flow cell before calibration. Perform calibration with the detectors or the Flow cells free from
contamination when the reading is stable.
<How to pour samples into the Flow cell>
● Use a funnel to pour the sample into the Flow cell. Pour it in the funnel.
● After pouring the sample into the Flow cell, drain the sample once. Pour another sample
again and measure the second sample.
● Do not allow bubbles to enter the sample. Pour the sample slowly down the side of the
funnel to help it free from bubbles as below.
● Do not allow the tip of the funnel to touch the sample in the cell to avoid any contamination
adhered to the surface of the funnel as below.
F7-10E.ai
Figure 7.1