<1. OVERVIEW>
1-4
IM 12E01B30-02EN 2nd Edition : June 09 , 2021-00
l
Head Tank
Name
Part No.
Description
Pressurized head tank for low turbidity K8003WA 2 NTU or less, same as option code /D1
Simple head tank
K8003YA over 2NTU to 10NTU, same as option code /D2
Head tank
K9658MR over 2 NTU, same as option code /D3
1.5
External Dimensions
See TB820D General Specification
1.6
Measurement principle
TB820D Turbidity detector uses the right angle scattered light method to continuously measure
the turbidity of your samples. When Sample is irradiated with light, the incident light is scattered
by the turbidity particle.
TB820D measures turbidity by using 660 nm or 860nm monochromatic light as the light source
and detecting light scattered in the sample in the 90-degree direction.
Scattered light intensity is calculated by light source intensity, turbidity and the structural
parameter of the unit. The equation is as follows.
Ls = K · Q · S
Where:
Ls: Scattered light intensity
K: Structural parameter (constant)
Q: light source intensity
S: Turbidity
TB820D continuously monitors the scattered light intensity (Ls) and light source intensity (Q) and
calculates the turbidity (S).