<8. Setting Parameters>
8-16
IM 01C27B01-01EN
8.3.16 Switching to Deep Sleep Mode
When the instrument will not be used for a long
time, switch the instrument to deep sleep mode to
conserve battery power. To switch to deep sleep
mode, follow the procedure below.
• Procedure to call up the switch-to-deep-sleep
parameter
TRANSDUCER block: Special Cmd
Set Deep Sleep mode(Standby) to the Special
Cmd parameter for the TRANSDUCER
block. To start from deep sleep mode, either
remove and insert the battery pack, or using
the provisioning device tool or the device
configuration tool via infrared communication.
In case of the external powered type, turn off
the power and on again.
CAUTION
After setting the deep sleep mode by infrared
device configuration tool, keep the infrared port
of device away from any other infrared signals.
NOTE
• Transmitter becomes the stop state after
setting deep sleep mode and cannot reply
any request from the device configuration
tool via wireless communication.
• For this reason, there is the case that an
error is display on the device configuration
tool via wireless communication.
• To wake up from deep sleep mode, please
pull battery pack and wait more than 30
seconds before attaching battery pack or
turning on the power.
8.3.17 Switching to Silence Mode
This is a function to pause the instrument when
it cannot join the field wireless network after a
specified time has elapsed. This function is effective
in conserving battery power when, for example, the
installation of the field wireless integrated gateway
is delayed compared to that of field wireless
devices. The default value is 28800 seconds (8
hours). Thereafter, a cycle of a one-hour pause and
six-minute search is repeated until the instrument
can join the field wireless network.
• Procedure to call up the switch-to-silence
parameter (Radio Silence)
UAPMO block: Radio Silence
Set 0 to 2
31
seconds for the Radio Silence
parameter of the UAPMO block. If 0 is set, the
Radio Silence parameter is invalid. To start from
deep sleep mode, either remove and insert the
battery pack, or use the provisioning device
tool or device configuration tool via infrared
communication.