13-15
IM 04L42B01-01E
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Remote Control (/R1)
Item
Specifications
Number of input terminals
8
Input type
Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common.
Input type and signal level
Voltage-free contact
Contact closed at 200 Ω or less and contact open at 100 kΩ or greater.
The remote contact input
operation can be set to normal open or normal close (release number 3 or later)
Open collector
ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 mA or less
Allowable input voltage
5 VDC
Signal type
Level or edge (250 ms or more)
Action
Executes a specified action by applying a given signal to the remote signal input terminal.
Action assignment: Set using the event action function
24 VDC Transmitter Power Supply (/TPS4 and /TPS8)
Item
Specifications
Number of loops
4 (/TPS4) or 8 (/TPS8)
Output voltage
22.8 to 25.2 VDC (under rated load current)
Rated output current
4 to 20 mADC
Max. output current
25 mADC (overcurrent protection operation current: approx. 68 mADC)
Allowable conductor resistance
RL ≤ (17.8 – minimum transmitter operation voltage)/0.02 A
where 17.8 V is the result obtained by subtracting the maximum drop voltage of 5 V when the
load shunt resistance is 250 Ω from the minimum output voltage of 22.8 V
Max. length of wiring
2 km (when using the CEV cable)
Insulation resistance
20 MΩ or more at 500 VDC between output terminal and ground
Dielectric strength
500 VAC (50/60 Hz, I = 10mA) for one minute between output terminal and ground
500 VAC (50/60 Hz, I = 10mA) for one minute between output terminals
Easy Text Entry (/KB1 and /KB2)
Item
Specifications
Remote control terminal (438227)
Operating temperature range
0 to 40°C
Operation humidity range
20% to 80% RH (at 5 to 40°C no condensation)
Power supply
3 VDC AA dry batteries × 2
Weight
Approx. 60 g (excluding the batteries)
External dimensions
170 (H) × 50 (W) × 23.7 (D) mm
Signal
Infrared
Combination with the DX
Number of units that can be controlled individually
Up to 32 units by setting the ID number
Communication distance
Up to 8 m from the front of the light receiving section of the DX (reference value)*
Directional characteristics
See the figure below (reference characteristics)*
* Varies depending on the operating environment such as the battery voltage and the presence
or absence of external light.
Standard accessories (/KB1)
Remote control unit, two alkaline dry batteries, and ID number label
Distance (m)
Horizontal
Vertical
0
0
0
30
15
15
10
10
5
5
1 1
-
30
-
45
-
60
45
60
90
-
90
Distance (m)
0
30
15
15
10
10
5
5
1 1
-
30
-
45
-
60
45
60
90
-
90
(+)
(+)
(-)
(-)
13.5 Options
Summary of Contents for Daqstation DX2000
Page 2: ......
Page 98: ...Blank...
Page 132: ...Blank...
Page 224: ...Blank...
Page 292: ...Blank...
Page 324: ...Blank...
Page 348: ...13 24 IM 04L42B01 01E 13 7 External Dimensions See the DX2000 Operation Guide IM04L42B01 02E...