1-12
IM 04L42B01-01E
• Updating of the Waveform
On the screen, 30 dots along the time axis is represented by a unit called division (see
the figure on the previous page). The displayed waveform is updated at an interval
corresponding to one dot. This interval is determined by the time corresponding to one
division (referred to as the trend interval). The relationship between the trend interval
and the speed of movement of waveforms on the screen is as follows:
Trend interval (/DIV)
5 s
*1
10 s
*1
15 s
*2
30 s
1 min
Time corresponding to one dot
(in seconds)
0.125
0.25
0.5
1
2
Speed of waveform movement
(approximation in mm/h)
10000
5000
2500
1250
625
Trend interval (/DIV)
2 min
5 min
10 min
15 min
20 min
Time corresponding to one dot
(in seconds)
4
10
20
30
40
Speed of waveform movement
(approximation in mm/h)
312
156
78
42
31
Trend interval (/DIV)
30 min
1 h
2 h
4 h
10 h
Time corresponding to one dot
(in seconds)
60
120
240
480
1200
Speed of waveform movement
(approximation in mm/h)
21
10
5.2
2.6
1.0
*1 40 dots per division. Selectable on the DX2004 and DX2008 (release number 3 or later).
*2 Selectable on the DX2010, DX2020, DX2030, DX2040, and DX2048 when the scan interval
is set to fast sampling mode (release number 3 or later).
Switching the Trend Interval
You can switch from the normal trend interval to the secondary trend interval during
memory sampling and vice versa. For the operating procedure, see section 5.3.
• Writing Messages
Preset messages
Free message
Set the message when writing the message
1
2
3
4
Start
Material 1
Trend display
Preset Messages
Preset messages are recalled and written.
The number of messages that you can use are 100 (message 1 to 10 are shared with
free messages). For the operating procedure, see section 5.4
Free Messages
Messages are entered when you need to enter them. The number of messages that
you can use are 10. For the operating procedure, see section 5.4.
Automatic Message Writing
• A message is written when the trend interval is switched during memory sampling.
For the setting procedure, see section 5.3.
• A message is written when the power recovers from a power failure during memory
sampling. For the operating procedure, see section 5.17.
• A message is written when the setting mode setup items are changed during
memory sampling (only on DXs with the /AS1 advanced security option).
1.3 Display
Summary of Contents for Daqstation DX2000
Page 2: ......
Page 98: ...Blank...
Page 132: ...Blank...
Page 224: ...Blank...
Page 292: ...Blank...
Page 324: ...Blank...
Page 348: ...13 24 IM 04L42B01 01E 13 7 External Dimensions See the DX2000 Operation Guide IM04L42B01 02E...
Page 366: ...Blank...