3-6
IM 04L42B01-01E
3.4 Setting the Moving Average of the Input
Set the moving average function of the measurement channel. This function suppresses
the effects of noise.
For a description of the function, see section 1.1.
Setup Screen
Press
MENU
(to switch to setting mode), and select the
Menu
tab >
Meas channel
>
Moving average
.
Setup Items
Average value
Scan Interval
Count
Average value
• First-CH/Last-CH
Select the target channels.
• On/Off
To use moving average, select
On
.
•
Count
Set the number of data points of the moving average in the range of 2 to 400.