5-1
IM CA700-02EN
1
2
3
4
5
6
App
Index
Specifications
5.1 Pressure Measurement Feature
Pressure Measurement Range*
Suffix Code
-01
02
-03
Pressure type
Gauge pressure
Measurement range
Positive pressure 0 to 200 kPa
Negative pressure –80 to 0 kPa
Positive pressure 0 to 1000 kPa
Negative pressure –80 to 0 kPa
Positive pressure 0 to 3500 kPa
Negative pressure –80 to 0 kPa
Measurement display range Up to 240.000 kPa
Up to 1200.00 kPa
Up to 4200.00 kPa
Measurement accuracy
(6 months)
after zero calibration
1, 2
Positive pressure
20 to 200 kPa: ±(0.01% of
r 0.003 kPa)
0 to 20 kPa: ±0.005 kPa
Positive pressure
±(0.01% of r 0.04 kPa)
Positive pressure
±(0.01% of r 0.15 kPa)
Negative pressure
±(0.2% of r 0.080 kPa)
Negative pressure
±(0.2% of r 0.08 kPa)
Negative pressure
±(0.2% of r 0.08 kPa)
Measurement accuracy
(1 year)
after zero calibration
1, 2
Positive pressure
20 to 200 kPa:±(0.01% of
r 0.010 kPa)
0 to 20 kPa: ±0.012 kPa
Positive pressure
±(0.01% of r 0.08 kPa)
Positive pressure
±(0.01% of r 0.30 kPa)
Negative pressure
±(0.2% of r 0.090 kPa)
Negative pressure
±(0.2% of r 0.09 kPa)
Negative pressure
±(0.2% of r 0.09 kPa)
Resolution
0.001 kPa
0.01 kPa
0.01 kPa
Response time
3
2.5 s or less
Allowable input
2.7 kPa abs to 500 kPa gauge 2.7 kPa abs to 3000 kPa gauge 2.7 kPa abs to 4500 kPa gauge
Internal volume
Approx. 6 cm
3
Temperature coefficient
4
±(0.001% of r 0.001% of full scale)/°C or less
Influence of positional setup Zero point ± 0.3 kPa or less
Measured medium
Gas and liquid (non-corrosive, non-flammable, non-explosive, and non-toxic fluids)
Measured medium
temperature
5
–10 to 50°C
Pressure sensor
Silicon resonant sensor
Pressure sensing element Diaphragm
Pressure unit display
kPa and other units (Pa, hPa, MPa, mbar, bar, atm, mmHg, inHg, gf/cm
2
, kgf/cm
2
, mmH
2
O@4°C,
mmH
2
O@20°C, ftH
2
O@4°C, ftH
2
O@20°C,inH
2
O@4°C, inH
2
O@20°C, torr, psi)
Input port
Rc 1/4 or 1/4 NPT female thread (selectable)
Measurement unit material Diaphragm: Hastelloy C276, Input port: SUS316
*
23°C ± 3°C
1
Zero-point calibration condition: Under atmospheric pressure
2
Yokogawa’s pressure standards accuracy is excluded.
3
Conditions of response time measurement:
The response time is defined as the time for the readout to settle within ±1% of the full scale from the
time when the positive pressure is released to atmosphere when it is at its full-scale value
(where the input unit is under no load).
4
Full scale of each model
CA700-E-01: Positive pressure 200 kPa and negative pressure 80 kPa
CA700-E-02: Positive pressure 1000 kPa and negative pressure 80 kPa
CA700-E-03: Positive pressure 3500 kPa and negative pressure 80 kPa
5
Liquid temperature 5 to 50°C
Pressure switch test
Holds the pressure valued at the time the port switches between open and close.
Measures the resistance between ports when they are closed (measurement current < 1 mA)
Resistance measurement range
0 to 2.000 kΩ
Tilt Alarm
The zero adjustment indicator blinks when the orientation of the CA700 changes by approximately
10 degrees or more after the previous zero adjustment.
If the external pressure sensor selected as a function changes in position of about 10 degrees or
more after zero adjustment, the zero adjustment indication blinks.
Chapter 5
Specifications
Summary of Contents for CA700
Page 14: ...xii...