Calibrating the Pressure T
ransmitter
3-9
IM CA700-02EN
1
2
3
4
5
6
App
Index
Setting the Calibration Direction (Direction)
Set the calibration direction.
17.
Press the ▲ or ▼ key to select
Direction
.
The options appear on the function menu.
0% to 100%
100% to 0%
0% to 100% to 0%
18.
Press
Up (F1)
,
Down (F2)
,
or Up/Down (F3)
.
If you select Up/Down, the number of calibration points will be set to (the number of points set
in step 16)x2 – 1.
Confirming the Settings
19.
After entering the information, press
Setting Done (F5)
. The settings are confirmed.
Press
ESC
to return the display.
Explanation
Calibration Procedure
You can register up to twenty sets of calibration processes. A calibration procedure consists of a set of
device information, measurement feature and source feature.
To calibrate a pressure device, select one of the registered processes to perform calibration.
Device Information
The device information that you set here is saved to a file along with the calibration data. This is
convenient in associating the saved data with the device.
Input Characters
The number of characters that you can enter is as follows.
Model No.: Up to 20 alphanumeric characters
Tag No.: Up to 32 alphanumeric characters
Serial No.: Up to 15 alphanumeric characters
3.2 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter