Calibrating the Pressure T
ransmitter
3-5
IM CA700-02EN
1
2
3
4
5
6
App
Index
Setting Measurement Conditions
Setting Measurement Features
5.
In the
calibration procedure setting display,
hold down the ▲ or ▼ key.
The display switches to page 2/3 (measurement condition setting display).
6.
Set the calibration target output.
Use the
▲ or ▼
key to select
Function
, and press
ENTER
.
The measurement feature options are displayed.
7.
Use the ▲ or ▼ key to select
mA LOOP
, and press
ENTER
.
ENTER
ENTER
Turns averaging on and off
Turns scaling on and off
Set the 0% value.
Set the 100% value.
Tolerance for pass/fail judgment
Confirms the settings
Set the unit
Setting Averaging and Scaling
Set averaging and scaling if necessary.
Here, we will set them to OFF.
8.
Use the ▲ or ▼ key to select
Averaging
, and press
ENTER
.
9.
Press
OFF (F2)
.
10.
Turn scaling off.
To use scaling, select
ON
, and press
ENTER
. A scaling setting menu appears.
3.2 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter