3-4
IM CA700-02EN
Setting the Device Information of the Calibration Target
Entering the Model Number (Model No.)
5.
Use the ▲ or ▼ key to select
Model No.
, and press
ENTER
.
A character input screen appears.
Moves the input
position
Backspace
Confirms the input
Switching the display from
uppercase l numbers input to
lowercase l symbols input
Input position
Selected character ( to move, ENTER to enter)
6.
Press (F1) or (F2) to move the input position, press , ,
▲,
or
▼
to select a character,
and press
ENTER
to enter the selected character.
7.
After entering the characters, press
Done (F5)
.
Entering the Tag No. and Serial No.
8.
Use the
▲
or
▼
key to select
Tag No.
or
Serial No.
, and press
ENTER
.
9.
Like the model number, enter the tag number and serial number.
Setting Other Items or Confirming the Settings
10.
To set measurement conditions and source conditions in addition to product information, press
▲ or ▼ key repeatedly until the screen changes. When the measurement condition or source
condition setup screen appears, continue with the procedure on the following pages.
To finish changing the settings, press
Setting Done (F5)
. The settings are confirmed.
3.2 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter