2-12
IM CA700-02EN
Connecting the External Pressure Seneor
WARNING
• Use the external pressure sensor correctly, following its operation manual. Improper use of
the external pressure sensor causes considerable danger because high pressure is applied
to it.
• Use only the pressure sensor specialized for the CA700.
French
AVERTISSEMENT
• Utilisez le capteur de pression externe correctement, en suivant son mode d’emploi.
L’utilisation incorrecte du capteur de pression externe entraîne un danger considérable car
une pression élevée est appliquée.
• Utilisez uniquement le capteur de pression spécialisé pour le CA700.
With the CA700 powered off, use the connection cable included with the external pressure sensor to
connect the external pressure sensor input terminal of the CA700 to the CA700 connection terminal of
the sensor.
Pressure transmitter
Pressure pump
(hand pump etc.)
CA700
External pressure sensor
External pressure sensor input terminal
Power on the CA700. When the external pressure sensor is recognized successfully, the icon
indicating that sensor appears on the screen of the CA700.
Note
• Be sure to power off the CA700 and then connect the external pressure sensor.
• The external pressure sensor is available in the following firmware versions of the CA700.
PM100-*-05 (16 MPa model): 1.10 or later
PM100-*-06 (70 MPa model): 1.20 or later
You can download the latest firmware version from our website.
• If you reconnect to an external pressure sensor with a different pressure range while the power of this
instrument is on, a message is displayed and this instrument is restarted.