3-8
IM CA700-02EN
12.
Turn scaling off.
To use scaling, select
ON
, and press
ENTER
. A scaling setting menu appears.
Setting the 0% Value and 100% Value
Set the pressure transmitter input pressure that correspond to the output current calibration range (0%
and 100%) of the pressure transmitter, which was set in steps 5 and 6.
13.
Use the ▲ or ▼ key to select
0% Value
, and press
ENTER
.
14.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set the 100% value.
ENTER
ENTER
Setting the Number of Calibration Points (Number Of Points)
15.
Use the ▲ or ▼ key to select
Number Of Points
, and press
ENTER
.
16.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
The range is 1 to 10.
ENTER
ENTER
3.2 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter