Setting Source Conditions
5.
In the
calibration procedure setting display,
hold down the ▲ or ▼ key.
The display switches to page 2/2 (source condition setting display).
Page 3/3
Setting the Source Feature
Here, we will set the source feature to P.
If you change the source feature to anything other than pressure (P or EXT P) here, the measurement
feature setting will change.
6.
Use the ▲ or ▼ key to select
Function
, and press
ENTER
.
The source function options are displayed.
7.
Use the
▲
or
▼
key to select P, and press
ENTER
.
The source function set.
Note
• Even if you set the source feature to pressure (P or EXT P), the CA700 will not produce pressure. The
CA700 will measure and display the pressure applied to the pressure transmitter.
• If the measurement feature is set to mA Loop when the source feature is set to anything other than
pressure (P or EXT P), the source feature is automatically set to pressure (P).
Setting the Unit
8.
Use the ▲ or ▼ key to select
Unit
, and press
ENTER
.
The pressure unit options are displayed.
9.
Use the
▲
or
▼
key to select a unit, and press
ENTER
.
The unit is set.
ENTER
ENTER
Setting Averaging and Scaling
The pressure transmitter input pressure can be averaged or scaled before it is displayed on the
CA700. Here, we will set both of them to OFF.
10.
Use the ▲ or ▼ key to select
Averaging
, and press
ENTER
.
11.
Press
OFF (F2)
.