33
4.
Operation Method
Setting/Display of Submenu
2. Setting/display of each item
2.3. Consumables replacement history
EXCH History
1. Hollow Filter
2. Ion-Exchange N
3. Ion-Exchange H
4. DW M-Filter
5. DI M-Filter
6. MHF
7.TOC-d UV Lamp
For display of the history of the pre-treatment
filter
・
ion exchange resin
(
N
)・
Ion exchange resin
(
H
)・
distilled water M filter
・
ion change water M filter
・
deaeration membrane
module
Cautions
1. During distillation and pure water sampling, the supply water circulation solenoid pump is always
running. When the distilled water tank is full, this pump stops. To sample pure water while this
pump is stopped, set this pump normally ON in the sub-menu and sample pure water after
stabilization of the water quality. This pump can be set to normal operation from the sub-menu while
the distillation water tank is full.
2. If the supply water circulation pump has been set to normally ON, its setting is cancelled under
following conditions:
①
Power key turned OFF
②
Distilled water sampled, the storage tank water level lowering, and distillation resumed
③
Abnormality
When the distilled wat
er tank is full, set the supply water circulation pump to “Normally ON” according to
Setting and Displaying Submenu, “2.8 PUMP manual operation” in Page 34.